Monte Carlo-based dose calculation for (32)P patch source for superficial brachytherapy applications.
Author | |
---|---|
Abstract |
:
Skin cancer treatment involving (32)P source is an easy, less expensive method of treatment limited to small and superficial lesions of approximately 1 mm deep. Bhabha Atomic Research Centre (BARC) has indigenously developed (32)P nafion-based patch source (1 cm × 1 cm) for treating skin cancer. For this source, the values of dose per unit activity at different depths including dose profiles in water are calculated using the EGSnrc-based Monte Carlo code system. For an initial activity of 1 Bq distributed in 1 cm(2) surface area of the source, the calculated central axis depth dose values are 3.62 × 10(-10) GyBq(-1) and 8.41 × 10(-11) GyBq(-1)at 0.0125 and 1 mm depths in water, respectively. Hence, the treatment time calculated for delivering therapeutic dose of 30 Gy at 1 mm depth along the central axis of the source involving 37 MBq activity is about 2.7 hrs. |
Year of Publication |
:
1969
|
Journal |
:
Journal of medical physics
|
Volume |
:
40
|
Issue |
:
1
|
Number of Pages |
:
13-7
|
Date Published |
:
1969
|
ISSN Number |
:
0971-6203
|
URL |
:
http://www.jmp.org.in/article.asp?issn=0971-6203;year=2015;volume=40;issue=1;spage=13;epage=17;aulast=Sahoo
|
DOI |
:
10.4103/0971-6203.152232
|
Short Title |
:
J Med Phys
|
Download citation |